DD6053 — Industrial gamma radioactivity detector
Rugged gamma detector in an IP54 aluminium housing, built around the C6020 measuring board. It measures gamma dose rate or surface contamination and transmits its results using the Dactyle protocol.
Components
- C6020 measuring board
- Detector holder and Geiger-Müller tube
- Communication interface and Dactyle connector
- High-strength IP54 aluminium housing
Functions
- Gamma dose rate or surface measurement, transmission via the Dactyle protocol
- Internal management: detector high-voltage supply, detection thresholds, LED status display, settings, programming mode
Communication
- Interfaces: RS-485, current loop, RS-232
- ASCII message exchange, 8 hardware addresses, coded commands and responses
- Commands: stop transmission, status request, advance tape, reset, send parameters, read signals, channel 1 / channel 2 results
- Responses: measurements, device status, parameters
- Protocols adaptable on request, with error recovery
Technical specifications
- Measuring range
- Gamma, from 10 µSv/h to 100 mSv/h (ZP1313 tube)
- Power supply
- 9–36 V or 1–5 V, consumption < 2 W
- Input / output
- Isolated digital, max. 100 V / 1 A
- Dimensions
- 190 × 112 × 61 mm
- Weight
- 720 g
- Housing
- Very strong aluminium, IP54
- Temperature
- Operating 0 to 60 °C — storage −20 to +80 °C